Polyethersulfone Hollow Fiber Hemodialyzer NF low-fow | ||||||||||||
Hemodialyzer model | NF-14L | NF-16L | NF-18L | NF-20L | ||||||||
Ultrafiltration coefficient (ml/hxmm Hg)±20% | 11 | 16 | 17 | 18 | ||||||||
Ground clearance QD=ml/min | ||||||||||||
QB (ml / min) | 200 | 300 | 400 | 200 | 300 | 400 | 200 | 300 | 400 | 200 | 300 | 400 |
Urea | 180 | 216 | 252 | 183 | 220 | 256 | 188 | 233 | 268 | 192 | 230 | 268 |
Creatinine | 170 | 204 | 221 | 175 | 210 | 227 | 180 | 222 | 239 | 183 | 219 | 237 |
Phosphate | 160 | 176 | 208 | 163 | 195 | 253 | 167 | 207 | 220 | 172 | 190 | 223 |
Vitamin B12 | 80 | 88 | 94 | 91 | 100 | 102 | 102 | 119 | 126 | 113 | 124 | 129 |
Ground clearance QD=800 ml/min | ||||||||||||
QB (ml / min) | 200 | 300 | 400 | 200 | 300 | 400 | 200 | 300 | 400 | 200 | 300 | 400 |
Urea | 183 | 222 | 262 | 188 | 226 | 264 | 188 | 233 | 268 | 196 | 236 | 273 |
Creatinine | 172 | 210 | 230 | 179 | 214 | 233 | 180 | 222 | 239 | 186 | 224 | 243 |
Phosphate | 163 | 180 | 216 | 166 | 199 | 259 | 167 | 207 | 220 | 175 | 194 | 227 |
Vitamin B12 | 84 | 93 | 100 | 94 | 103 | 112 | 102 | 119 | 126 | 117 | 132 | 132 |
Surface area (m2) not less | 1.4 | 1.6 | 1.8 | 2.0 | ||||||||
Blood flow range(mL/min) | 200~400 | |||||||||||
Dialysis flow range(mL/min) | 500~800 | |||||||||||
Blood cavity hydraulic resistance (QB=300ml/min) | ≤8 | ≤8 | ≤8 | ≤8 | ||||||||
Filling volume (ml)±10% | 90 | 90 | 90 | 90 | ||||||||
Maximum pressure of use | 500 mmHg Art |